ML-based muon identification using a FNAL-NICADD scintillator chamber for the MID subsystem of ALICE 3

T0 review · 4 major / 5 minor · reviewed 2026-08-06 · deepseek-v4-flash

Pith's one-line read A scintillator muon chamber with 30:1 pion rejection validates the ALICE 3 MID baseline.

desk verdict Useful, honest test-beam R&D for ALICE3 MID, but the headline pion-suppression number is conditional on a tuned GEANT4 simulation that is not validated at the classifier level. read the letter →

arxiv 2507.02817 v2 pith:5LMWOCKJ submitted 2025-07-03 physics.ins-det hep-ex

classification physics.ins-dethep-ex
keywords muonidentificationplasticscintillatorsiliconphotomultiplierboosteddecisiontreespionsuppressiontestbeamALICE3MIDabsorberthickness
verification ladder T0 review T1 audit T2 compute T3 formal

The pith

A machine-rendered reading of the paper's core claim, the machinery that carries it, and where it could break.

The reading

The paper reports test-beam results for a two-layer scintillator muon chamber intended for the ALICE 3 muon-identifier (MID) subsystem. Using 3 GeV/c pion- and muon-enriched beams and an iron absorber of variable thickness, it claims that a machine-learning classifier (boosted decision trees) trained on simulated hit position, time-over-threshold, and time-of-arrival keeps 98% of muons while suppressing pions by a factor of about 30 at the reference 70 cm absorber thickness. The machine-learning selection improves hadron rejection by about 15% over a simple time-over-threshold cut. If correct, the baseline MID design of plastic scintillator bars with wavelength-shifting fibers and silicon photomultipliers behind a roughly four-interaction-length iron absorber can deliver the J/psi dimuon tagging the ALICE 3 physics programme needs in Run 5.

What carries the argument

The load-bearing mechanism is a boosted decision tree trained on eight per-layer quantities: the hit positions x and y, the time-over-threshold (ToT), and the time-of-arrival (ToA) in the two orthogonal scintillator layers, which together separate muon signal from pion background. Its inputs come from a Monte Carlo simulation of the prototype in which the photon transport is parametrized, namely an 8% photon collection efficiency, ToT that is linear up to 72 photoelectrons, and Gaussian ToA smearing fitted to the measured distributions. The classifier cutoff is fixed at the value that yields 98% muon efficiency per bar, and the simulation also encodes the tuned beam composition; applying the trained classifier to statistically independent samples of $10^{6}$ particles gives the quoted pion efficiencies.

What would settle it

Run the same prototype at the same 3 GeV/c momentum behind a 70 cm iron absorber with the beam particle type identified event-by-event by an independent device, such as a threshold Cherenkov counter or a calorimeter that separates muons, pions, and electrons, then apply the published classifier cut and count the true pion fraction. If the pion-candidate efficiency for a pure pion beam comes out above about 3.6% at the 98%-muon-efficiency working point, the claimed suppression factor of 30 would be contradicted.

Watch

Extended reading notes

Core claim

The central experimental claim is that the measured pion-candidate efficiency with the trained boosted decision tree is about 3.0 ± 0.15% for a 70 cm iron absorber when the muon efficiency is set to 98% in a single bar, which corresponds to a pion-suppression factor of ≈30 ± 1.5; the suppression rises to ≈50 ± 2.5 at 80 cm and ≈100 ± 5 at 90 cm. The paper further claims that the machine-learning analysis improves hadron rejection by approximately 15% relative to a traditional time-over-threshold-only cut, and that this measurement, cross-checked with an independent multiwire proportional chamber analysis, is consistent with ALICE 3 MID requirements. A secondary result is the inferred beam composition: the pion-enriched beam is about 68% pions, 29.5% electrons, and 2.5% muons, while the muon-enriched beam has a purity of 78.0 ± 1.6%.

Load-bearing premise

The result stands on the assumption that the machine-learning classifier trained on Monte Carlo simulated detector response performs the same on real data; the simulation's photon collection, time-over-threshold linearity, time-of-arrival smearing, and beam composition are all tuned until the simulation matches the measured distributions, so any mismatch in the tails those variables use for muon or pion separation would bias the quoted efficiency and suppression factor without necessarily showing up in the raw distributions.

Editorial extensions

If this is right

  • At the reference 70 cm absorber, the MID baseline keeps 98% of muons per bar with a pion suppression factor of about 30; adding 10 cm of iron raises the suppression to about 50, and 20 cm raises it to about 100.
  • Switching from a ToT-only discriminator to the boosted decision tree buys roughly 15% more hadron rejection with no hardware change, so the same absorber thickness can tolerate more background.
  • The measured pion efficiency of about 3% at 70 cm indicates that the scintillator-plus-absorber combination meets the ALICE 3 MID requirement for competitive J/psi signal-to-background in both pp and Pb-Pb collisions.
  • The beam-composition analysis implies that simulations of this test beam must include roughly 30% electrons in the negative pion beam; ignoring this electron contamination misrepresents the background seen in the chamber.
  • Geometrical acceptance of 85 to 87% for 3 GeV/c muons is limited by the 3.5 mm bar spacing, and the paper's planned reduction of dead areas would raise the acceptance above 95%.

Reading between the lines

Editorial extensions of the paper, not claims the author makes directly.

  • A likely extension is to train the same classifier with hit multiplicity and inter-layer coincidence time on a full-size 1 m prototype; the paper lists those as future variables, and one would expect the suppression factor at a fixed absorber thickness to grow beyond the quoted values.
  • The roughly 30% electron contamination found in the pion-enriched beam suggests that other test-beam studies using this beam line should treat the electron component explicitly, otherwise pion-rejection measurements could be diluted by electrons that deposit energy in the scintillator like muons.
  • The 15% gain from machine learning over a simple ToT cut implies that the hardware requirement on time resolution could be relaxed if the classifier is used, trading a precise time measurement for a cheaper analogue readout; the paper does not quantify that trade.
  • A testable consequence of the exponential absorber-length trend is that the same classifier should transfer to other absorber materials; a comparative run with a different material would show whether the suppression factor scales with nuclear interaction length as the observed trend suggests.
Share X Bluesky LinkedIn Reddit HN

Editorial analysis

A structured set of objections, weighed in public.

Desk editor's note, referee report, and a circularity audit.

Referee Report

4 major / 5 minor

Summary. The paper reports a test-beam study of a two-layer FNAL-NICADD extruded-scintillator prototype chamber with WLS-fiber/SiPM readout, intended for the ALICE 3 MID subsystem. Data were taken at the CERN PS T10 beamline with 3 GeV/c pion- and muon-enriched beams and iron absorbers of 60, 70, 80, 90, and 100 cm. The authors train a BDT in GEANT4 on position, ToT, and ToA variables, apply it to data, and quote a pion-candidate efficiency of about 3.0% at 70 cm, corresponding to a pion-suppression factor of about 30 for 98% single-bar muon efficiency. They also compare the BDT result with a traditional ToT cut, report a roughly 15% improvement in hadron rejection, and infer beam-composition parameters (68% pions, 29.5% electrons, 2.5% muons for the pion-enriched beam; 78% muon purity for the muon-enriched beam) by matching GEANT4 to measured distributions.

Significance. If the central suppression claim is valid, the result is valuable for the ALICE 3 MID baseline: it demonstrates that a relatively simple scintillator-bar chamber plus an ML selector can meet the required pion rejection at 70 cm absorber thickness, and it quantifies the gain over a traditional ToT cut. The paper also provides useful practical information on prototype construction, trigger setup, and the T10 beam composition. The independent MWPC-based cross-check is a positive feature, and the simulation-vs-data comparisons cover all absorber thicknesses used in the scan. However, the significance is currently conditional: the quoted suppression factor depends on a BDT trained exclusively on a GEANT4 simulation whose detector-response parameters and beam composition are tuned to the same data, and the validation is limited to one-dimensional marginals. The strength of the claim therefore rests on an unquantified simulation-to-data transfer of the classifier, which is the main point that needs to be addressed.

major comments (4)
  1. [Sec. 4 and Figs. 4, 5, 9] The central pion-suppression claim rests on a BDT trained entirely on GEANT4 simulations whose detector response is tuned to data: the 8% photoelectron-collection parameter, the ToT-versus-p.e. linearity up to 72 p.e., the Gaussian ToA smearing, and, in Sec. 5, the beam composition (68% pions, 29.5% electrons, 2.5% muons) are all adjusted until MC matches data. The validation shown in Figs. 4 and 5 is limited to one-dimensional ToT, ToA, and position distributions, which do not constrain the correlations among x, y, ToT, and ToA that the BDT exploits. No comparison of the BDT output distribution and no classifier-level closure test are presented, so the quoted pion-candidate efficiency of 3.0% and the suppression factor of about 30 inherit an unquantified simulation-to-data transfer error. I recommend adding a data/MC comparison of the BDT response variable, or a data-trained cross-check using the muon-enriched sample to calibrate the signal response, and assigning an explicit systematic for the transfer.
  2. [Sec. 5 and Figs. 7, 9] It is not clear whether the number quoted as the measured pion efficiency (3.0 +/- 0.15% at 70 cm) is obtained directly from data with the BDT applied, or taken from the beam-composition-matched GEANT4 simulation. The text states that the pion-enriched data are described by a simulated beam of 68% pions, 29.5% electrons, and 2.5% muons, while the 'pure pion beam' efficiency shown in the middle panel of Fig. 7 is a GEANT4-only result. Because the beam composition is a fitted parameter, a data-derived efficiency depends on the assumed contamination; the paper should state explicitly which curve in Fig. 9 is the data measurement and should propagate the beam-composition uncertainty into the quoted 3.0 +/- 0.15% value.
  3. [Sec. 5 and Fig. 9] The systematic uncertainty of about 5% assigned from the difference between the MID-chamber and MWPC analyses covers analysis choices but not the dominant systematic from the MC-to-data transfer of the BDT or from the tuned simulation parameters (ToA smearing, ToT response, 8% photon collection). The quoted uncertainty of +/- 0.15% on 3.0% therefore appears to reflect only a small part of the total uncertainty budget. The authors should either justify that the MWPC comparison captures the classifier-transfer uncertainty, or enlarge the systematic to include variations of the tuned simulation parameters and the beam composition within their allowed ranges.
  4. [Sec. 5 and Fig. 7] The claimed 15% improvement of the BDT over the traditional ToT cut is evaluated only in GEANT4 simulation, as the left and middle panels of Fig. 7 are both labeled as GEANT4 results. This improvement therefore inherits the same unvalidated transfer from simulation to data. A data-based comparison of the two selectors, at least for the pion-enriched beam, would be needed to support the statement that the ML selection improves hadron rejection by about 15% in the real detector.
minor comments (5)
  1. [Fig. 1 caption] The caption contains a typo: 'Tigger 1' should read 'Trigger 1'.
  2. [Fig. 3] The y-axis labels in Fig. 3 appear garbled (for example, 's etadidnac' should be 'candidates'); the figure should be regenerated with correct labels.
  3. [Sec. 4] The text says that a momentum resolution of 1% was assumed and that a 20% resolution was also studied with stable results, but no distributions or numbers from that stability study are shown; a brief summary or reference would help support the statement.
  4. [Sec. 5] The sentence 'Both of them were corrected for geometrical acceptance' should specify how the acceptance correction is applied to data and to simulation, since the two are compared in Fig. 9.
  5. [Sec. 5] The text says an electron contamination of 25% to 30% is possible and then chooses 29.5% in the simulated composition; the reason for this specific value within the quoted range should be stated more explicitly.

Circularity Check

2 steps flagged · score 4.0 of 10

Two fitted calibration inputs (muon beam purity and pion-beam composition) are presented as measured results, partially feeding the quoted pion-suppression factor; no equation-level circularity and the data comparison retains independent test-beam content.

  1. fitted input called prediction [Section 5, final paragraph (caption of Fig. 9); Sec. 5, middle panel of Fig. 7]
    "A pion-enriched beam consisting of 68% pions, 29.5% electrons and 2.5% muons was simulated, and our data is found to be consistent with this assumption within one sigma. Based on the discussion above, and given the agreement between the data and Monte Carlo simulations, the measured pion efficiency is around 3.0±0.15% (70cm absorber thickness)"

    The 'measured pion efficiency' of 3.0% is the GEANT4 pure-pion output (middle panel of Fig. 7: 'the pion efficiency amounts to around 3% using the trained BDT'), not a value derived from data alone. The data are only 'consistent' with a simulation whose beam composition (68% pions, 29.5% electrons, 2.5% muons) was selected within an external range to achieve that consistency, while both the original 98% pion/2% muon composition and the new composition are said to describe the data. Thus the agreement does not independently pin down the pure-pion efficiency; presenting the simulation value as 'measured' and then quoting the Sec. 6 suppression factor (≈30) makes that factor a calibrated simulation output rather than an independent datum.

  2. fitted input called prediction [Section 5, final paragraph, after Fig. 9]
    "MC simulations assuming a muon purity of 78% describe the data. Our result suggests that the muon-enriched beam delivered at the PS T10 has a purity of 78.0±1.6% for momentum of 3GeV/c."

    The 78% purity is the simulation input that is varied until the Monte Carlo muon-candidate efficiency matches the measured absorber-length dependence; the same number is then quoted as the determined beam property with a fitted uncertainty. The agreement with data is therefore largely a result of choosing this parameter, and the 'result suggests' is the fitted input. The independent MWPC analysis does provide a 2% systematic cross-check, and the efficiency-vs-thickness shape gives some constraint, so the step is not a pure tautology, but it is a fitted parameter reported as a measurement.

full rationale

The paper's central result—a pion-suppression factor of ≈30 at 70 cm absorber—is obtained by training a BDT on GEANT4 and applying it to test-beam data. The simulation is not ab initio: Sec. 4 fixes the photon-collection efficiency at 8%, makes ToT linear up to 72 photoelectrons, and uses a Gaussian fitted to the measured ToA distributions to smear times. Sec. 5 then adopts beam compositions (68% pions/29.5% electrons/2.5% muons for the pion-enriched beam; 78% muon purity for the muon-enriched beam) to make the Monte Carlo match the measured candidate efficiencies. The final 'measured pion efficiency' of 3.0±0.15% is the GEANT4 pure-pion value (Fig. 7, middle), not a data-only extraction, and the quoted suppression factor is that simulation ratio together with the 98% muon-efficiency cut. This is a genuine partial circularity: the simulation output is relabeled as a measurement after calibrating its inputs to the same data. However, it is not a full constructional tautology. The measured pion-candidate efficiency points in Fig. 9 are real data and could have disagreed with the simulation; the consistency is asserted for a composition range suggested by an external technical note; and the independent MWPC-based analysis provides a 5% systematic on the pion-candidate efficiency and a 2% systematic on the muon purity. These elements give the central comparison independent empirical content beyond the fitted parameters. No load-bearing self-citation chain or uniqueness-argument circularity was found; the cited prior work [6] is a separate test-beam characterization and is not the basis of the suppression claim. The main limitation—namely that only one-dimensional ToT/ToA/position marginals are compared, while the BDT may use their correlations—is a simulation-to-data transfer risk, not itself a circularity under the stated rules. Score 4 reflects the two fitted-input-as-result steps and the partial overlap between the simulation calibration and the reported performance numbers.

Assumptions & free parameters 4 free parameters · 3 assumptions · 0 invented entities

The key result inherits its calibration from GEANT4 simulations whose detector-response parameters and beam composition are tuned to match the same data being analyzed; the central suppression factor is not obtained from a standalone data measurement. No new physical entities are introduced.

free parameters (4)
  • pion-enriched beam composition = 68% pions, 29.5% electrons, 2.5% muons
    Chosen so that GEANT4 reproduces the measured pion-candidate efficiency curve in Fig. 9; this composition directly affects the inferred true pion efficiency and suppression factor.
  • muon-enriched beam purity = 78.0 ± 1.6%
    Fitted by comparing MC muon-candidate efficiency with data; reported as the beam purity, but it is a simulation input matched to data rather than an independent measurement.
  • ToA smearing parameters (Gaussian mean and width per channel) = not quoted
    A Gaussian function was fitted to the measured ToA distributions and used to smear simulated times (Sec. 4); the simulation must match data for the BDT to transfer.
  • ToT versus photoelectron response parameters = linear up to 72 p.e., reduced slope beyond; 8% photon collection
    Chosen by hand to mimic fiber trapping and SiPM response in GEANT4 (Sec. 4); these tune the simulated ToT shapes to the measured data.
assumptions (3)
  • domain assumption GEANT4 correctly simulates hadronic shower development and muon energy loss through 60-100 cm iron at 3 GeV/c.
    The central suppression numbers assume the MC transport is accurate; no external validation is provided beyond the visual beam-profile agreement in Fig. 2.
  • domain assumption A BDT trained on GEANT4 output applies without bias to the real prototype.
    If classifier inputs (ToT, ToA, positions) from simulation differ from data in the tails used for discrimination, the measured pion-candidate efficiency will be biased. Agreement is checked by eye, not with a quantitative classifier-level test.
  • ad hoc to paper The T10 beam composition model from Ref. [12] (25-30% electron contamination) applies to this run.
    The paper adopts this technical document's electron fraction to resolve a mismatch between the original 98% pion assumption and the data; the applicability to this specific 3 GeV/c run is assumed.

how reviews work

0 comments
Cite this review

Pith. "Pith review of ML-based muon identification using a FNAL-NICADD scintillator chamber for the MID subsystem of ALICE 3." pith.science (2026). https://pith.science/paper/5LMWOCKJ

@misc{pith2026250702817,
  author       = {Pith},
  title        = {Pith review of: ML-based muon identification using a FNAL-NICADD scintillator chamber for the MID subsystem of ALICE 3},
  year         = {2026},
  howpublished = {\url{https://pith.science/paper/5LMWOCKJ}},
  note         = {Machine review of arXiv:2507.02817}
}
abstract

The ALICE Collaboration is planning to construct a new detector (ALICE 3) aiming at exploiting the potential of the high-luminosity Large Hadron Collider (LHC). The new detector will allow ALICE to participate in LHC Run 5 scheduled from 2036 to 2041. The muon-identifier subsystem (MID) is part of the ALICE 3 reference detector layout. The MID will consist of a standard magnetic iron absorber ($\approx4$ nuclear interaction lengths) followed by muon chambers. The baseline option for the MID chambers considers plastic scintillation bars equipped with wave-length shifting fibers and readout with silicon photomultipliers. This paper reports on the performance of a MID chamber prototype using 3 GeV/$c$ pion- and muon-enriched beams delivered by the CERN Proton Synchrotron (PS). The prototype was built using extruded plastic scintillator produced by FNAL-NICADD (Fermi National Accelerator Laboratory - Northern Illinois Center for Accelerator and Detector Development). The prototype was experimentally evaluated using varying absorber thicknesses (60, 70, 80, 90, and 100 cm) to assess its performance. The analysis was performed using Machine Learning techniques and the performance was validated with GEANT 4 simulations. Potential improvements in both hardware and data analysis are discussed.

Figures

Figures reproduced from arXiv: 2507.02817 by the authors.

Figure 1
Figure 1. Experimental setup. The trigger scintillators are labeled as Tigger 1, Trigger 2, Trigger 3, and Trigger 4. The beam [PITH_FULL_IMAGE:figures/full_fig_p004_1.png] view at source ↗
Figure 2
Figure 2. The beam profiles for 3 GeV/c muon- and pion-enriched beams are displayed in the left- and right-hand side plots, respectively. An iron absorber 70 cm thick is considered. GEANT 4 simulations assumed a pure muon beam, whereas the pion beam considers a 2% muon contamination. See the text for details behind this assumption. 4. Monte Carlo simulations and data analysis The simulation of the detector geometry and the pr… view at source ↗
Figure 3
Figure 3. (Left) Hit position distribution in the first layer of the MID chamber simulated with GEANT 4, particles are generated [PITH_FULL_IMAGE:figures/full_fig_p005_3.png] view at source ↗
Figures from the paper (6 more)
Figure 4
Figure 4. Figure 4: Time-over-threshold distributions measured in the 10 channels of the prototype. The upper panel show the results for the [PITH_FULL_IMAGE:figures/full_fig_p006_4.png]
Figure 5
Figure 5. Figure 5: Time-of-arrival distributions measured in the 10 channels of the prototype. The upper panel show the results for the first [PITH_FULL_IMAGE:figures/full_fig_p006_5.png]
Figure 6
Figure 6. Figure 6: The variables used for training the Boosted Decision Trees for the muon tagging in the MID prototype. The position [PITH_FULL_IMAGE:figures/full_fig_p008_6.png]
Figure 7
Figure 7. Figure 7: (Left) Absorber length dependence of the fraction of muon candidates tagged with BDT in the pion-enriched beam [PITH_FULL_IMAGE:figures/full_fig_p009_7.png]
Figure 8
Figure 8. Figure 8: Pion-candidate efficiency as a function of absorber thickness using a 3 GeV/ [PITH_FULL_IMAGE:figures/full_fig_p010_8.png]
Figure 9
Figure 9. Figure 9: Pion-candidate (left) and muon-candidate (right) efficiency as a function of absorber length. Data are compared with [PITH_FULL_IMAGE:figures/full_fig_p011_9.png]

Discussion (0). Sign in to comment.

Reference graph

Works this paper leans on

12 extracted references · 6 canonical work pages

  1. [1]

    The ALICE experiment – A journey through QCD (11 2022).arXiv:2211.04384. 10 50 60 70 80 90 100 110 (cm) LAbsorber length, 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 0.1 -candidate efficiency π pion-enriched beam, 2024 MID test beam c3 GeV/data (FNAL-NICADD scintillator chamber) ) -μ, 2.5% - , 29.5% e-πGEANT 4 (68% ) -μ, 2.0% -πGEANT 4 (98% 50 60 70 80 9...

  2. [2]

    R. Bala, I. Bautista, J. Bielcikova, A. Ortiz, Heavy-ion physics at the LHC: Review of Run I results, Int. J. Mod. Phys. E 25 (07) (2016) 1642006.arXiv:1605.03939,doi:10.1142/S0218301316420064

  3. [3]

    Letter of intent for ALICE 3: A next-generation heavy-ion experiment at the LHC, CERN-LHCC-2022-009, LHCC-I-038 (11 2022).arXiv:2211.02491

  4. [4]

    Dainese, A

    A. Dainese, A. Di Mauro, Scoping document for the ALICE 3 detector, Tech. rep., CERN, Geneva (2025). URLhttps://cds.cern.ch/record/2925455

  5. [5]

    Pla-Dalmau, A

    A. Pla-Dalmau, A. D. Bross, K. L. Mellott, Low-cost extruded plastic scintillator, Nucl. Instrum. Meth. A 466 (2001) 482–491.doi:10.1016/S0168-9002(01)00177-2

  6. [6]

    R. Alfaro, et al., Characterisation of plastic scintillator paddles and lightweight MWPCs for the MID subsystem of ALICE 3, JINST 19 (04) (2024) T04006.arXiv:2401.04630,doi:10.1088/1748-0221/19/ 04/T04006

  7. [7]

    Radiation damage of SiPMs

    E. Garutti, Y. Musienko, Radiation damage of SiPMs, Nucl. Instrum. Meth. A 926 (2019) 69–84.arXiv: 1809.06361,doi:10.1016/j.nima.2018.10.191

  8. [8]

    Ahdida, D

    C. Ahdida, D. Bozzato, D. Calzolari, F. Cerutti, N. Charitonidis, A. Cimmino, A. Coronetti, G. L. D’Alessandro, A. Donadon Servelle, L. S. Esposito, R. Froeschl, R. García Alía, A. Gerbershagen, S. Gilar- doni, D. Horváth, G. Hugo, A. Infantino, V. Kouskoura, A. Lechner, B. Lefebvre, G. Lerner, M. Magistris, A. Manousos, G. Moryc, F. Ogallar Ruiz, F. Pozz...

Show all 12 references
  1. [9]

    Varga, G

    D. Varga, G. Kiss, G. Hamar, G. Bencedi, Close cathode chamber: Low material budget MWPC, Nucl. Instrum. Meth. A 698 (2013) 11–18.doi:https://doi.org/10.1016/j.nima.2012.09.025. URLhttps://www.sciencedirect.com/science/article/pii/S0168900212010728

  2. [10]

    Agostinelli, et al., GEANT4 - A Simulation Toolkit, Nucl

    S. Agostinelli, et al., GEANT4 - A Simulation Toolkit, Nucl. Instrum. Meth. A 506 (2003) 250–303. doi:10.1016/S0168-9002(03)01368-8

  3. [11]

    Hocker, P

    A. Hocker, P. Speckmayer, J. Stelzer, J. Therhaag, E. von Toerne, H. Voss, M. Backes, T. Carli, O. Cohen, A. Christov, D. Dannheim, K. Danielowski, S. Henrot-Versille, M. Jachowski, K. Kraszewski, J. Krasz- nahorkay, A., M. Kruk, Y. Mahalalel, R. Ospanov, X. Prudent, A. Robert...

  4. [12]

    van Dijk, A

    M. van Dijk, A. Hayat, D. Banerjee, J. Bernhard, B. Gokturk, L. Nevay, J. Petersen, M. Schwinzerl, Particle production and identification for the T10 secondary beamline of the CERN East Area (7 2025). arXiv:2507.02567. 12

Pith tools

Reviewed August 6, 2026 · model on record in the stance chip above.