A novel active quenching circuit for single photon detection with Geiger mode avalanche photodiodes
Reviewed by Pithpith:G7M5CPMOopen to challenge →
classification
quant-ph
keywords
circuitphotonsingleactivedetectionquenchingavalanchedetector
read the original abstract
In this paper we present a novel construction of an active quenching circuit intended for single photon detection. For purpose of evaluation, we have combined this circuit with a standard avalanche photodiode C30902S to form a single photon detector. A series of measurements, presented here, show that this single photon detector has a dead time of less than 40ns, maximum random counting frequency of over 14MHz, low after pulsing, detection efficiency of over 20% and a good noise performance. This simple and robust active quenching circuit can be built from of-the-shelf electronic components and needs no complicated adjustments.