Magnetic shield of PMT used in DAMPE electromagnetic calorimeter

classification ⚛️ physics.ins-det
keywords magneticfieldgainr5610a-01ablealongaxiscalorimeter
0
0 comments X
read the original abstract

The magnetic characteristics of photomultiplier tube R5610A-01 are studied in this paper. The experimental data shows that the gain of R5610A-01 loses about 53% when the magnetic field is 3G along its +X axis. A cylinder of one-layer permalloy strip is able to reduce the effect of 3G magnetic field on the PMT's gain to less than 1%.