The SELEX Phototube RICH Detector

classification ✦ hep-ex physics.ins-det
keywords detectorrichringselexarticlebetacentralcherenkov
0
0 comments X
read the original abstract

In this article, construction, operation, and performance of the RICH detector of Fermilab experiment 781 (SELEX) are described. The detector utilizes a matrix of 2848 phototubes for the photocathode to detect Cherenkov photons generated in a 10m Neon radiator. For the central region an N0 of 104/cm, corresponding to 13.6 hits on a beta=1 ring, was obtained. The ring radius resolution measured is 1.6%.