A Time-Of-Flight-Based Reconstruction for Real-Time Prompt-Gamma Imaging in Protontherapy

Not yet reviewed by Pith; the record is open.

This paper has not been read by Pith yet. Machine review is queued; the pith claim, tier, and objections will appear here once it completes.

SPECIMEN: schema-true, not a live event

T0 review · schema-true

One-sentence machine reading of the paper's core claim.

pith:XXXXXXXX · record.json · timestamp

arxiv 2012.09275 v1 pith:DCJLBLTV submitted 2020-12-16 physics.med-ph

classification physics.med-ph
keywords beamprotonincidentprotonsrangereconstructionresolutionshift
verification ladder T0 review T1 audit T2 compute T3 formal
0 comments
abstract

We are currently conceiving, through (MC) simulation, a multi-channel gamma detector array (TIARA for Time-of-flight Imaging ARrAy) for the online monitoring of protontherapy treatments. By measuring the Time-Of-Flight (TOF) between a beam monitor placed upstream and the Prompt-Gamma (PG) detector, our goal is to reconstruct the PG vertex distribution to detect a possible deviation of proton beam delivery. In this paper, two non-iterative reconstruction strategies are proposed. The first is based on the resolution of an analytical formula describing the PG vertex distribution in 3D. Here, it was resolved under a one-dimensional approximation in order to measure a potential proton range shift along the beam direction. The second is based on the calculation of the Centre-Of-Gravity (COG) of the TIARA pixel detectors counts and also provides 3D information on a possible beam displacement. The PG vertex reconstruction was evaluated in two different scenarios. A coincidence time resolution of 100 ps (rms) can be attained in single proton regime (operating a reduction of the beam current) and using an external beam monitor to provide a start trigger for the TOF measurement. Under these conditions, MC simulations have shown that a millimetric proton range shift sensitivity can be achieved at 2$\sigma$ with 10$^{8}$ incident protons. This level of accuracy would allow to act in real-time if the treatment does not conform to treatment plan. A worst case scenario of a 1 ns (rms) TOF resolution was also considered to demonstrate that a degraded timing information can be compensated by increasing the acquisition statistics: in this case, a 2 mm range shift would be detectable at 2$\sigma$ with 10$^{9}$ incident protons. The COG method has shown excellent capabilities of detecting lateral beam displacements: a 2 mm sensitivity was found at 2$\sigma$ with 10$^{8}$ incident protons.

Discussion (0). Continue with ORCID to comment.

Forward citations

Cited by 1 Pith paper

Reviewed papers in the Pith corpus that reference this work. Sorted by Pith novelty score. Full citation record

  1. Real-Time Requirements and Transferability in Compton Imaging: From the Detector Chain to the Application

    physics.med-ph 2026-08 accept novelty 7.0 of 10

    A marker-based review of 83 Compton imaging papers shows the field reports single operating points and omits latency, cost and source-complexity terms, and that deployment commitment tracks gap-filling and handover co...

Pith tools