arxiv: 1604.02600 · v1 · pith:5HIQW3R2new · submitted 2016-04-09 · 🌌 astro-ph.IM · physics.ins-det

A multiplexer for the ac/dc characterization of TES based bolometers and microcalorimeters

classification 🌌 astro-ph.IM physics.ins-det
keywords bolometerspowerarraymicrocalorimetersmultiplexersronx-rayarrays
0
0 comments X
read the original abstract

At SRON we are developing the Frequency Domain Multiplexing (FDM) for the read-out of the TES-based detector array for the future infrared and X-ray space mission. We describe the performances of a multiplexer designed to increase the experimental throughput in the characterisation of ultra-low noise equivalent power (NEP) TES bolometers and high energy resolving power X-ray microcalorimeters arrays under ac and dc bias. We discuss the results obtained using the TiAu TES bolometers array fabricated at SRON with measured dark NEP below $5\cdot 10^{-19}W/Hz^{1/2}$ and saturation power of several fW