pith. sign in

arxiv: 1303.5204 · v2 · pith:YBQPZWATnew · submitted 2013-03-21 · ❄️ cond-mat.mes-hall

Optimizing atomic resolution of force microscopy in ambient conditions

classification ❄️ cond-mat.mes-hall
keywords atomicambientcantileverdampinghydrationlayeramplitudeconditions
0
0 comments X
read the original abstract

Ambient operation poses a challenge to AFM because in contrast to operation in vacuum or liquid environments, the cantilever dynamics change dramatically from oscillating in air to oscillating in a hydration layer when probing the sample. We demonstrate atomic resolution by imaging of the KBr(001) surface in ambient conditions by frequency-modulation atomic force microscopy with a cantilever based on a quartz tuning fork (qPlus sensor) and analyze both long- and short-range contributions to the damping. The thickness of the hydration layer increases with relative humidity, thus varying humidity enables us to study the in uence of the hydration layer thickness on cantilever damping. Starting with measurements of damping versus amplitude, we analyzed the signal and the noise characteristics at the atomic scale. We then determined the optimal amplitude which enabled us to acquire high-quality atomically resolved images.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.