A low-cost sEMG wristband prototype passed safety tests with leakage currents of 11.4-13.5 uA and showed signal correlation r > 0.85 with a commercial reference system under a standards-based validation protocol.
Title resolution pending
1 Pith paper cite this work. Polarity classification is still indexing.
1
Pith paper citing it
fields
eess.SP 1years
2026 1verdicts
UNVERDICTED 1representative citing papers
citing papers explorer
-
Design of a validation methodology for a prototype wristband for capturing muscle signals and upper limb movement
A low-cost sEMG wristband prototype passed safety tests with leakage currents of 11.4-13.5 uA and showed signal correlation r > 0.85 with a commercial reference system under a standards-based validation protocol.