Stability comparison of two absolute gravimeters: optical versus atomic interferometers

classification ⚛️ physics.atom-ph physics.geo-phphysics.ins-det
keywords comparisonabsolutegravimetersatomicbetterdifferentluxembourgoptical
0
0 comments X
read the original abstract

We report the direct comparison between the stabilities of two mobile absolute gravimeters of different technology: the LNE-SYRTE Cold Atom Gravimeter and FG5X\#216 of the Universit\'e du Luxembourg. These instruments rely on two different principles of operation: atomic and optical interferometry. The comparison took place in the Walferdange Underground Laboratory for Geodynamics in Luxembourg, at the beginning of the last International Comparison of Absolute Gravimeters, ICAG-2013. We analyse a 2h10 duration common measurement, and find that the CAG shows better immunity with respect to changes in the level of vibration noise, as well as a slightly better short term stability.