pith. sign in

arxiv: 1808.07468 · v1 · pith:GZ6HYFX5new · submitted 2018-08-21 · ⚛️ physics.ins-det · physics.optics

Cavity attenuated phase shift Faraday rotation spectroscopy

classification ⚛️ physics.ins-det physics.optics
keywords cavityrotationattenuateddetectionfaradayphasesensitivityshift
0
0 comments X
read the original abstract

Cavity attenuated phase shift Faraday rotation spectroscopy has been developed and demonstrated by oxygen detection near 762 nm. The system incorporates a high-finesse cavity together with phase-sensitive balanced polarimetric detection for sensitivity enhancement and achieves a minimum detectable polarization rotation angle (1{\sigma}) of $5.6\times 10^{-9}$ rad/$\surd$Hz, which corresponds to an absorption sensitivity of $4.5\times 10^{-10}$ cm$^{-1}/{\surd}$Hz without the need for high sampling rate data acquisition.