arxiv: 2302.06979 · v1 · pith:VFGADIU4 · submitted 2023-02-14 · physics.med-ph

Characterization of breast tissues in density and effective atomic number basis via spectral X-ray computed tomography

Reviewed by Pith T0 review T1 audit T2 compute T3 formal T4 kernel pith:VFGADIU4record.jsonopen to challenge →

classification physics.med-ph
keywords tissuesatomiceffectivenumberspectralbreastdensityenergy
0
0 comments X
read the original abstract

Differentiation of breast tissues is challenging in X-ray imaging because tissues might share similar or even the same linear attenuation coefficients $\mu$. Spectral computed tomography (CT) allows for more quantitative characterization in terms of tissue density and effective atomic number by exploiting the energy dependence of $\mu$. In this work, 5 mastectomy samples and a phantom with inserts mimicking breast soft tissues were evaluated in a retrospective study. The samples were imaged at three monochromatic energy levels in the range of 24 - 38 keV at 5 mGy per scan using a propagation-based phase-contrast setup at SYRMEP beamline at the Italian national synchrotron Elettra. A custom-made algorithm incorporating CT reconstructions of an arbitrary number of spectral energy channels was developed to extract the density and effective atomic number of adipose, fibro-glandular, pure glandular, tumor, and skin from regions selected by a radiologist. Preliminary results suggest that, via spectral CT, it is possible to enhance tissue differentiation. It was found that adipose, fibro-glandular and tumorous tissues have average effective atomic numbers (5.94 $\pm$ 0.09, 7.03 $\pm$ 0.012, and 7.40 $\pm$ 0.10) and densities (0.90 $\pm$ 0.02, 0.96 $\pm$ 0.02, and 1.07 $\pm$ 0.03 g/cm$^{3}$) and can be better distinguished if both quantitative values are observed together.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.