pith. sign in

arxiv: physics/0611290 · v1 · submitted 2006-11-29 · ⚛️ physics.med-ph · physics.ins-det

Multi-Channel SQUID System for MEG and Ultra-Low-Field MRI

classification ⚛️ physics.med-ph physics.ins-det
keywords systemfieldimagingmagneticmulti-channelresolutionsquidultra-low-field
0
0 comments X
read the original abstract

A seven-channel system capable of performing both magnetoencephalography (MEG) and ultra-low-field magnetic resonance imaging (ULF MRI) is described. The system consists of seven second-order SQUID gradiometers with 37 mm diameter and 60 mm baseline, having magnetic field resolution of 1.2-2.8 fT/rtHz. It also includes four sets of coils for 2-D Fourier imaging with pre-polarization. The system's MEG performance was demonstrated by measurements of auditory evoked response. The system was also used to obtain a multi-channel 2-D image of a whole human hand at the measurement field of 46 microtesla with 3 by 3 mm resolution.