The reviewed record of science sign in
Pith

arxiv: 2402.05983 · v1 · pith:55ZZLLHS · submitted 2024-02-08 · eess.IV · cs.LG· physics.app-ph· physics.ins-det

Capability enhancement of the X-ray micro-tomography system via ML-assisted approaches

Reviewed by Pith T0 review T1 audit T2 compute T3 formal T4 kernel pith:55ZZLLHSrecord.jsonopen to challenge →

classification eess.IV cs.LGphysics.app-phphysics.ins-det
keywords artifactsimagesringx-rayanalysiscausesdetectormicro-ct
0
0 comments X
read the original abstract

Ring artifacts in X-ray micro-CT images are one of the primary causes of concern in their accurate visual interpretation and quantitative analysis. The geometry of X-ray micro-CT scanners is similar to the medical CT machines, except the sample is rotated with a stationary source and detector. The ring artifacts are caused by a defect or non-linear responses in detector pixels during the MicroCT data acquisition. Artifacts in MicroCT images can often be so severe that the images are no longer useful for further analysis. Therefore, it is essential to comprehend the causes of artifacts and potential solutions to maximize image quality. This article presents a convolution neural network (CNN)-based Deep Learning (DL) model inspired by UNet with a series of encoder and decoder units with skip connections for removal of ring artifacts. The proposed architecture has been evaluated using the Structural Similarity Index Measure (SSIM) and Mean Squared Error (MSE). Additionally, the results are compared with conventional filter-based non-ML techniques and are found to be better than the latter.

This paper has not been read by Pith yet.

discussion (0)

Sign in with ORCID, Apple, or X to comment. Anyone can read and Pith papers without signing in.